Skip to main content
. 2009 Mar 19;9(3):2076–2087. doi: 10.3390/s90302076

Table 2.

Performance of REFETs with PVC membranes with different contents of DNP.

DNP/PVC + DNP Drift (mV/h) Sensitivity (mV/pH) Linearity (%)
50 wt. % −0.70 Unstable w/o
60 wt. % −0.74 10.4 ± 2.2 99.7 ± 0.3
70 wt. % −1.05 9.2 ± 1.2 97.6 ± 1.3
80 wt. % −0.21 35.0 ± 2.9 97.4 ± 1.1