Table 2.
Comparison of computational and Doppler assessment of the proximal flow events. Simulation and measurement of the 3D PISA and VC area for a circular (0.39cm2) and rectangular (0.35cm2) orifice under different flow conditions. CFD PISA is compared to direct measurement of 3D color Doppler surface area at defined aliasing velocities. CFD simulated and 3D color Doppler VC area were compared to the actual orifice area (OA), since known. Vena Contracta, VC. Proximal Isovelocity Surface Area, PISA. Aliasing velocity, Va (cm/s)
| PISA | VC AREA | ||||||
|---|---|---|---|---|---|---|---|
| 3D color Doppler | CFD Simulation | Relative Difference | 3D color Doppler | rel. diff. with OA | CFD Simulation | rel. diff. with OA | |
| Circular Orifice | |||||||
| 30 ml | 2.0cm2(Va=36) | 2.4cm2(Va=36) | 20% | 0.40cm2 | 3% | 0.36cm2 | 8% |
| 70 ml | 3.0cm2(Va=41) | 3.7cm2(Va=41) | 23% | 0.40cm2 | 3% | 0.36cm2 | 8% |
| Rectangular Orifice | |||||||
| 30 ml | 2.7cm2(Va=36) | 2.9cm2(Va=36) | 7% | 0.40cm2 | 14% | 0.32cm2 | 9% |
| 70 ml | 3.7cm2(Va=36) | 4.1cm2(Va=36) | 11% | 0.40cm2 | 14% | 0.32cm2 | 9% |
| Combined Mean | 15.3% | 8.5% | |||||