Table 3.
Troubleshooting guide for microscope parameters
Problem | Possible Cause | Possible Remedy |
---|---|---|
Specimen moves while imaging | Poor contact with mounting substrate | Press specimen more firmly on surface of mount |
Poor contact of specimen to mount | Apply conductive paint around specimen to improve contact | |
Excessive charging of sample | Inadequate bulk conductivity | Decrease number of electrons entering sample or Increase number of electrons exiting sample (See Supplemental Table 1) |
Specimen oxidized | ||
Image appears stretched | Astigmatism from contaminants | Stigmate aperture |
Unable to stigmate | Misaligned objective aperture | Center beam and aperture |