Published in final edited form as: Phys Med Biol. 2012 Apr 5;57(9):2477–2490. doi: 10.1088/0031-9155/57/9/2477

Table 3.

Linear attenuation coefficients (LACs) measured from the bovine femur phantom from experiment 1. Errors from CT-based measures compared to the 511 keV PET transmission image are also given. The coefficient of variation of all LAC values ranged from 3% to under 1%.

Energy 511 keV 140 kVp 120 kVp 80 kVp




Material & region LACTX (cm−1) LACCTAC (cm−1) ΔLAC (%) LACCTAC (cm−1) ΔLAC (%) LACCTAC (cm−1) ΔLAC (%)
Compact bone 1* 0.151 0.149 −1.3 0.141 −6.8 0.129 −14.1
Compact bone 2 0.150 0.148 −1.3 0.140 −6.6 0.129 −13.6
Cancellous bone 1 0.096 0.093 −3.1 0.093 −2.9 0.092 −3.4
Cancellous bone 2 0.096 0.094 −2.5 0.093 −2.6 0.093 −3.0
Contrast cylinder 0.097 0.124 27.3 0.123 26.7 0.124 27.5
*

Numbers correspond with regions indicated in Figures 2 and 4.