Skip to main content
. 2012 May 2;2012:571201. doi: 10.1100/2012/571201

Table 2.

LC-MS/MS parameters (MRM transition, DP, EP, CEP, CE, and CXP) and retention times of four immunosuppressive drugs and internal standards.

CyA TRL SRL RAD Ascomycin IS CyD IS
MRM-transition 1202.80 → 425.20 821.60 → 768.50 931.66 → 864.50 975.65 → 908.50 809.59 → 756.40 1216.90 → 425.50
DP (V) 100.00 50.63 41.25 44.38 50.63 110.00
EP (V) 10 10 10 10 10 10
CEP (V) 39.85 34.88 38.52 39.97 34.49 47.93
CE (V) 68 29 23 250 29 69
CXP (V) 4 12 14 14 12 4
Retention time (min) 1.97 1.81 1.84 1.85 1.81 2.00
Retention time RSD% (min) 0.245 0.137 0.186 0.142 0.148 0.236

CyA: cyclosporin A, TRL: tacrolimus, SRL: sirolimus, RAD: everolimus, CyD: cyclosporin D, IS: internal Standard, MRM: multiple reaction monitoring, DP: declustering potential, EP: entrance potential, CEP: collision cell entrance potential, CE: collision energy, CXP: collision cell exit potential, and V: volt.