Skip to main content
. 2012 Mar 27;12(4):4133–4155. doi: 10.3390/s120404133

Table 2.

Comparison of the different calibration techniques analyzed.

Parameters Method

Line Dujon DBSE (8bits)

Number of fibers located. Initial RT entries 49,127 49,127 49,127
Final validated entries 46,454 (94.5%) 40,241 (81.9%) 42,711 (86.9%)
Corrected entries (redundant) 3,270 1,920 5,867
Eliminated entries 2,672 (5.4%) 6,077 (14.1%) 6,416 (13%)
Mean scan time 7.91 min 2.6 min 5.54 min
Mean RT calculation time 38.94 min 1.3 min 2.2 min
Mean analysis time of redundancies and outliers 13.98 min 2.5 min 5.36 min
Number of images used 522 * 16 36
Final image size [pixels] 261 × 2611 254 × 254 254 × 254
*

A scan space of 261 × 261 images in each dimension was considered. This inflated size of the grid is subsequently corrected in the TR so that the size of the image is not greater than nfibmax = 256 in each dimension, eliminating those cell positions that do not have an appreciable influence on the fibers.