Table 1.

Characteristics of dengue rapid diagnostic tests mentioned in this paper.

Manufacturer Product name Analytes Storage temperature (°C) Quoted accuracy (Sn/Spa) Sampleb 10/20 c Formatd Sample volume (ul) Maximum time (min)e
Merlin Dengue Fever IgG and IgM Combo Device IgM/IgG 2–30°C IgM 96/98
IgG 97/98
S/P/WB Yes LF 1 30
Standard Diagnostics BIOLINE Dengue Duo NS1 antigen and IgG and IgM Combo Device NS1 Ag
IgM/IgG
1–30°C NS1-Ag 92.8/98.4
IgM/IgG 99.4/93.0
S/P/WB Yes LF NS1-Ag 100
IgM/IgG 10
20
Biosynex Immunoquick Dengue Fever IgG and IgM IgM/IgG 2–30°C IgM 97.6/98.3
IgG 95.2/96.6
S/P/WB Yes W 1 20
Biorad STRIP NS1 Ag 2–8°C NS1-Ag 92.3/98.8 S/P No W 50 15
Alere Panbio Dengue Early Rapid Kit NS1 Ag 2–8°C Not stated S No LF 50 15
Alere Panbio Dengue Duo Cassette IgM/IgG 2–8°C Sb convalescent—10c−85.1/91.6; 20−98.8/91.6
P acute—10−58.3/45.0; 20−100/45.0
WB acute—10−71.4/91.2; 20−77.4/91.2
WB convalescent—10−78.6/85.3; 20−100/85.3
S/P/WB Yes LF 10 15
MP Diagnostics  ASSURE IgA 2–28°C Not stated S/P/WB No LF 25 15–20

Sn/spa: sensitivity/specificity.

bS–serum; P: plasma; WB: whole blood.

cPrimary and secondary infections: manufacturer claims RDT can differentiate

dW: wickstyle; LF: lateral flow.

eMaximum time in minutes to confirm a negative result.