Skip to main content
. Author manuscript; available in PMC: 2012 May 23.
Published in final edited form as: Nat Protoc. 2010 Jun;5(6):993–1004. doi: 10.1038/nprot.2010.48

TABLE 1.

Troubleshooting table.

Step Problem Possible reason Solution
7 No peaks (oligosaccharides) are observed in SAX- or RPIP-HPLC and UV detection Enzyme is inactive Check for enzyme activity on a GAG standard
Insufficient enzymatic treatment Extend the digestion time
Baseline chromatogram in SAX- or RPIP-HPLC and UV detection is noisy and contains spurious peaks Solvent inadequately degassed Use an on-line vacuum degasser to the HPLC system
Purity of reagents has been compromised Reprepare fresh reagents and HPLC solvents
GAG sample was impure Repurify GAG sample and treat again with lyase
12 Overlapping of oligosaccharides species in RPIP-HPLC-ESI-MS pH values of HPLC solvents are different than 7.0 Carefully check for pH of HPLC solvents. Reprepare fresh HPLC solvents
Sample contains salt Either repeat purification and depolymerization of GAG sample or desalt oligosaccharides
Elution time periods change or become unstable from run to run in RPIP-HPLC-ESI-MS The column has not been well conditioned from run to run Reequilibrate the C18 column(s) in solvent C for 20 min at a flow rate of 0.5 ml min−1 (for the Discovery C18 column) or 0.3 ml min−1 (for the Gemini C18 column)
No peaks are evident by mass spectrometry Solvent flow is inadequate Make sure the solvent is flowing from the needle
Make sure that the divert valve setting is correct
High voltages and spay chamber currents are off Make sure the electrospray high voltages are switched on
Drying gas flow is insufficient Check the drying gas flow and temperature
Pressure is too low Make sure the fore and high-vacuum pressure are within normal ranges