Skip to main content
. 2012 May 21;6:761–769. doi: 10.2147/OPTH.S31690

Table 1.

Mean central corneal thickness taken with Topcon SP-3000P noncontact specular microscope, ultrasound pachymeter DGH-500, and Artemis-2 very high-frequency ultrasound

Device Mean CCT μm (±SD) Minimum Maximum 95% CI of mean
Session 1
SP-3000P 520 ± 29 466 567 509–530
DGH-500 560 ± 33 507 613 548–571
Artemis-2 558 ± 33 502 619 546–569
Session 2
SP-3000P 520 ± 30 453 572 508–531
USP 561 ± 37 486 621 548–574
Artemis-2 554 ± 34 506 616 542–566

Abbreviations: CCT, central corneal thickness; CI, confidence interval; SD, standard deviation; USP, ultrasound pachymeter.