Table 2.

Bland and Altman analysis of central corneal thickness measurements obtained by SP-3000P, Artemis-2, and DGH-500 ultrasound pachymetry

Techniques Mean difference (μm) SD 95% CI (μm) P value CoR
SP-3000Pa −0.9 4.7 −10.0 to 8.2 >0.05 ± 9.1
SP-3000Pb −0.9 5.5 −11.6 to 9.9 ± 10.8
Artemis-2a −0.5 4.1 −8.6 to 7.6 >0.05 ± 8.1
Artemis-2b −0.9 4.6 −9.9 to 8.1 ± 9.0
DGH-500a 0.6 6.3 −11.8 to 13.0 >0.05 ± 12.4
DGH-500b −1.7 6.2 −13.9 to 10.4 ± 12.1

Notes: Mean difference within session, standard deviations, 95% confidence intervals (mean ± 1.96 SD) in each technique with their CoR* are represented. P value is the between-observer paired t-test analysis.

a

SP-3000P is first session measurement;

b

SP-3000P is second session measurement;

*

statistically significant at the 0.05.

Abbreviations: CI, confidence interval; CoR, coefficients of repeatability; SD, standard deviation.