TABLE IV.

Reconstruction Sharpness and, Noise Level Measurements for the Physical Phantom Study

Motion Phase Motion rate (bpm) View-angle sampling (view angles) Pseudo PSF width (mm)
Pseudo PSF shift (mm)
Noise standard deviation (10–4mm–1)
Vertical Horizontal Vertical Horizontal
A 75 17 0.45 1.02 0.45 0.60 2.2
B 75 17 0.40 0.36 0.49 0.50 2.0
C 75 17 0.51 1.1 0.48 0.68 2.4
D 75 16 0.50 0.95 0.42 0.53 1.9

Assuming an acquisition time of 14 seconds.

The ROI used for the rRMSE measurements is shown in Fig. 9.

Both the prior image and the reference images have noise standard deviations of 4.6 × 10–4 mm–1.