Skip to main content
. 2011 Dec 13;470(7):1932–1940. doi: 10.1007/s11999-011-2213-3

Table 4.

Comparisons of our mechanical fixation failure rates with those of fixed-bearing cementless cups

Study Grist-blasted cups % of mechanical fixation failures Porous-coated cups % of mechanical fixation failures Mean followup (years) or intervals of followup
Reikeras and Gunderson [28] 25% 7–10
Lai et al. [16] 38% 10
Miyakawa et al. [24] 31% 11–14
Kim et al. [15] 14% 7
Malchau et al. [20] 18% 6
Bohm and Bosche [3] 0% 11
Manley et al. [21] 11% 2% 7.9
D’Lima et al. [7] 6% 6
Bojescul et al. [4] 17% 15
Engh et al. [11] 0,4% 7–15
Garavaglia et al. [13] 0% 10
Vicente et al. [29] 12.3% 11.4% 10–12
Current study 3.7% 1% 7.7