Skip to main content
. 2012 Feb 8;470(7):1860–1868. doi: 10.1007/s11999-012-2248-0

Fig. 5.

Fig. 5

Through-thickness total wear penetration rate versus in vivo duration is shown for the Sigma® fixed bearing inserts, including those in rough Ti trays with 4-Mrad gamma-irradiated and barrier-packaged material (Ti 4 MR GB) and those in polished CoCr trays with (1) 4-Mrad gamma-irradiated and barrier-packaged material (CoCr 4 MR GB) and (2) 5-Mrad gamma-irradiated and gas plasma-sterilized material (CoCr 5 MR XL). Fixed bearings showed an increasing wear rate with increasing in vivo duration.