Skip to main content
. 2011 Nov 11;112(7):2539–2547. doi: 10.1007/s00421-011-2230-7

Table 3.

Validity results, showing the mean ± SD values of physiological variables measured by the three metabolic systems during parallel data collection at rest, moderate, and vigorous cycle exercise

Activity Douglas bag MetaMax 3B Oxycon Pro
V E (BTPS, L min−1) Rest 8.72 ± 1.03 7.93 ± 1.73 7.72 ± 1.82
Moderate 29.89 ± 8.19 30.61 ± 9.78^ 28.73 ± 8.31
Vigorous 66.00 ± 20.52 66.25 ± 19.16 62.54 ± 18.26
VO2 (STPD, L min−1) Rest 0.27 ± 0.04 0.30 ± 0.08^ 0.25 ± 0.06
Moderate 1.12 ± 0.33 1.24 ± 0.41*^ 1.12 ± 0.36
Vigorous 2.32 ± 0.46 2.59 ± 0.52*^ 2.37 ± 0.44
VCO2 (STPD, L min−1) Rest 0.22 ± 0.04 0.26 ± 0.07^ 0.20 ± 0.06
Moderate 1.03 ± 0.31 1.17 ± 0.40*^ 1.01 ± 0.33
Vigorous 2.25 ± 0.52 2.64 ± 0.66*^ 2.17 ± 0.43

* Significantly different to Douglas bag measurements (ANOVA, p < 0.05)

^ Significantly different to Oxycon Pro measurements (ANOVA, p < 0.05)