Table 3.

Dimensions of the fabricated photosensitive-polyimide-based multichannel neural electrodes.

Description ECoG electrode Mesh intracortical electrode Guide cnnula electrode
Design dimensions [μm] Fabrication dimensions [μm] Shrinkage/expansion effect compared to design dimensions [%] Design dimensions [μm] Fabrication dimensions [μm] Shrinkage/expansion effect compared to design dimensions [%] Design dimensions [μm] Fabrication dimensions [μm] Shrinkage/expansion effect compared to design dimensions [%]
Number of recording microelectrodes 16 16 5 5
Diameter of recording microelectrode 100 108.5 ± 1.5 8.5 40 43.7 ± 0.1 9.3
Pitch of recording microelectrode site 500 (dorso-ventral) 487.6 ± 1.3 −2.5 80 74.5 ± 0.6 −6.9
700 (rostral-caudal) 685.7 ± 1.6 −2.0
Interconnection line 25 24.9 ± 0.8 −0.4 15 (mesh area) 10.3 ± 1.2 −31.3
15 (probe area) 14.3 ± 0.3 −4.7
Minimum space between interconnection lines 25 24.5 ± 0.6 −2.0 35 (mesh area) 39.8 ± 0.9 13.7
35 (probe area) 34.5 ± 0.4 −1.4
Probe width 1500 1499.50 0.0 310 321.7 ± 0.7 3.8
Insulated probe widthin mesh area 30 (the most left side) 28.8 ± 0.6 −4.0
30 (mesh area) 22.6 ± 0.6 −24.7
Minimum space between insulated probes in mesh area 20 27.8 ± 0.7 39.0
Length of short-side of through-holes 110 116.4 ± 2.3 5.8
Length of long-side of through-holes 300 307.5 ± 4.7 2.5