Skip to main content
. 2012 May-Jun;46(3):259–265. doi: 10.4103/0019-5413.96369

Figure 3.

Figure 3

X-rays of cadaveric femoral specimen (a) with unstable reverse oblique trochanteric fracture with posteromedial defect implanted with 95° dynamic hip screw and it was subjected to cyclical compression load. (b) It failed after 4482 cycles. Plate bending of 15°, plate barrel junction deformation of 2.5°, and screw bending of 1.5° were observed at the end of cyclical loading