Table 3.

The Experimental Results of Rise Time, Overshoot, and Steady-State Errors by Examining the Variation of One MRAC Control Variable at a Time for Adaptive MRI Thermal Feedback Control of Ultrasound Hyperthermia

Control Parameters Experimental Results Summary


Exp.
No.
Time Constant
(Min)
Adaptation
Gain
Initial Values Rise Time
(Min)
Overshoot
(°C)
Steady-State
Error (°C)
1 = 3 0.005 Optimized value 3 ± 0.5 1.9 ± 0.6 1.8 ± 0.5
= 8 0.005 Optimized value 8 ± 0.5 1.3 ± 0.3 1.5 ± 0.6
2 4τ = 7.8 0.005 Optimized value 8.1 ± 0.5 1.3 ± 0.3 1.5 ± 0.5
4τ = 7.8 0.015 Optimized value 7.85 ± 0.5 1.5 ± 0.4 1.5 ± 0.5
3 4τ = 6 0.005 Zero 12 ± 0.5 5.2 ± 0.6 3.5 ± 0.5
4τ = 6 0.005 Optimized value 6 ± 0.5 1.1 ± 0.3 1.5 ± 0.6

For each experiment (i.e., 1, 2, or 3), the control parameter under investigation is marked as bold.