Table 3.

Insulin Dose Determinations by the Subjectsa

All subjects Meter Total

Correct Incorrect
Manual method Correct 145 (35%) 8 (2%) 153 (37%)

Incorrect 241 (59%) 15 (4%) 256 (63%)

Total 386 (94%) 23 (6%) 409
Cohort 1 (easy) Meter Total

Correct Incorrect
Manual method Correct 64 (31%) 2 (1%) 66 (32%)

Incorrect 136 (65%) 6 (3%) 142 (68%)

Total 200 (96%) 8 (4%) 208
Cohort 2 (advanced) Meter Total

Correct Incorrect
Manual method Correct 81 (40%) 6 (3%) 87 (43%)

Incorrect 105 (52%) 9 (4%) 114 (57%)

Total 186 (93%) 15 (7%) 201
a

Each subject determined two insulin doses by manual calculation and two insulin doses by using the FreeStyle InsuLinx meter. Cohort 1 (104 subjects) used the easy formula in manual calculation and the easy mode on the meter. Cohort 2 (101 subjects) used the advanced formula in manual calculation and the advanced mode on the meter.