Skip to main content
. 2012 May 15;17(8):081402. doi: 10.1117/1.JBO.17.8.081402

Fig. 5.

Fig. 5

(a) The source-detector layout of the Gen-1 hand-held probe, with all fibers spaced 0.5 cm apart. The typical output intensity from each of the six simultaneous illumination sources (solid red circles) is shown. The mean and standard deviation of output source light intensity is 1.02±0.937  mW, making it a nonhomogenous source distribution. (b) The source-detector layout of each of the two probe heads of the Gen-2 imager. The typical output intensity from each of the (three each) simultaneous illumination sources (solid red circles) is shown. The mean and standard deviation of output source light intensity for the two probe heads is 2.01±0.03  mW and 2.09±0.06  mW, making it a homogenous source distribution.