Table 4.

Assessment of accuracy and precision of sunitinib and N-desethyl sunitinib in plasmaa

Nominal Concentration (ng/mL) Calculated Concentration (ng/mL) Accuracy (%) Bias Precision (%)
RSD (%) Between Run Recovery (%)
Within Run Between Run
Sunitinib
0.1 (LLOQ) 0.092 ±0.003 91.94 −8.06 3.08 3.31 3.31 -b
0.3 0.306 ± 0.021 102.11 2.11 5.66 7.04 7.04 93.81
8 8.20 ± 0.49 102.55 2.55 2.74 6.01 6.01 97.17
80 80.6 ± 4.7 100.78 0.78 1.88 5.86 5.86 97.39
80.0 (1:10)c 81.2 ± 3.9 101.50 1.5 4.16 4.79 4.79 -b
N-desethyl sunitinib
0.2 (LLOQ) 0.191 ± 0.011 95.50 −4.5 5.64 5.90 5.90 -b
0.6 0.609 ± 0.042 101.58 1.58 6.49 6.92 6.92 90.34
16.0 16.6 ± 0.7 103.50 3.5 3.11 4.32 4.32 101.1
160.0 154.0 ± 5.7 96.25 −3.75 3.02 3.69 3.69 98.13
160.0 (1:10)c 160.8 ± 9.2 100.50 0.5 5.08 5.75 5.75 -b
a

Performed in quintuplicate on three separate days

b

Not determined

c

Sample diluted 1:10 (v/v) prior to analysis