Table 13: Technical Failures and Procedural Complications Reported in 5 Health Technology Assessments.

Type of Problem Taylor et
al. 2005
(77)
Mailis-Gagnon
et al. 2004 (51)
Cameron
2004 (49)
Turner et al.
2004 (19)
Grabow et
al. 2003 (59)
Lead problems 27% 4% Migration: 9.7%
Breakage: 13.2%
23.1% 8.3%–42.8%
Generator-related problems 6% None reported 1.6 5.8 None reported
Extension cable problems 10% None reported Not reported None reported None reported
Reoperation None reported None reported None reported 23.1% 11.1%–50%
Subcutaneous dissection of generator pocket None reported 8.3% None reported None reported None reported
Infection 6% 4% 3.4% 4.5% superficial

0.1% deep
1.4%–11.7%
CSF* leak 7% Not reported 0.3% Not reported Not reported
*

CSF indicates cerebrospinal fluid.