Table 6: Prediction Probabilities of Bispectral Index Monitors With 4 Anesthetic Techniques (29).
Prediction probability (Pk)) | ||
---|---|---|
Anesthetic Technique | Mean | SEM* |
1: Sevoflurane/low-dose remifentanil | 0.684 | 0.61 |
2: Sevoflurane/high-dose remifentanil | 0.668 | 0.061 |
3: Propofol/low-dose remifentanil | 0.743 | 0.056 |
4: Propofol/high-dose remifentanil | 0.721 | 0.057 |
Combined | 0.685 | 0.029 |
SEM indicates standard error of the mean.