Table 1.

The reconstructed volume (in cc) of the targets imaged with CBCT in static and dynamic modes obtained using auto-contouring for WW −500 to 0 and the percentage difference in volume with reference CT respectively.

Targets Reconstructed Volume using auto-contouring technique for WW −500 to 0 (in cc)
Reference CT volume CBCT (static) Volume difference (%) CBCT (dynamic PA1) Volume difference (%) CBCT (dynamic PA2) Volume difference (%)
Dumbbell 43.71 42.37 −3.07 31.94 −26.93 25.94 −40.65
Cylinder 42.17 41.61 −1.32 36.12 −14.35 33.19 −21.29
U-shape 38.19 35.92 −5.94 27.82 −27.15 22.02 −42.34
Solid triangle 26.53 25.01 −5.73 18.32 −30.95 14.91 −43.80
Sphere 20.68 20.13 −2.66 15.65 −24.32 12.24 −40.81