Skip to main content
. Author manuscript; available in PMC: 2012 Jul 10.
Published in final edited form as: IEEE Trans Neural Syst Rehabil Eng. 2009 Mar 16;17(2):146–155. doi: 10.1109/TNSRE.2009.2017805

Figure 5. Participant sEMG during Serial Speech.

Figure 5

The difference between words and pauses in the percent above baseline RMS sEMG from each recording location during serial speech using a traditional EL (labelled as “Initial”) and during use of the EMG-EL with visual feedback (labelled as “During Feedback”) for each participant. Electrode positions for which each participant achieved at least 80% serial speech performance using the EMG-EL are indicated with background shading in grey.