Table 1.

Validation parameters and acceptance criteria

Validation parameter Acceptance criterion Remark
Specificity Nonblocked/blocked ratio ≥ 4
Matrix effect Effect of matrix ≤ 30% Criteria adjusted to cell-based assay performance
Linearity 0.7 < 90% CI of slope < 1.3
Repeatability CV ≤ 30% Criteria adjusted to cell-based assay performance
Intermediate precision CV ≤ 30% Criteria adjusted to cell-based assay performance
LOQ For information purposes only
LOD For information purposes only
Accuracy Difference between HRIG and CL184 ≤ 30% Accuracy is measured as concordance between HRIG and CL184
Stability Difference between stability sample and comparator sample ≤ 30% Criteria adjusted to cell-based assay performance