Table 1.

Areas of tendon damage caused by anchors.

Sample number (a) Healix Peek anchor (5.5 mm) (b) Fastin RC anchor (5.0 mm) (c) Bio-Corkscrew Suture anchor (5.0 mm) (d) Healix Transtend anchor inserted through the Percannula system (4.0 mm) (e) Healix Transtend anchor (3.4 mm) without cannula
1 25.0 17.4 23.3 7.0 13.0
2 28.6 22.6 21.4 10.7 10.0
3 31.6 22.8 22.9 9.1 10.1
4 37.5 18.6 24.7 10.9 16.7
5 25.8 22.7 22.9 6.6 18.7
6 26.8 18.3 23.7 7.5 14.1
7 28.5 20.2 24.8 11.7 13.1
Mean 29.1 20.4 23.4 9.1 13.7
SD 4.3 2.3 1.2 2.1 3.2
P (versus a)  0.0005 0.0052 0.0000 0.0000
P (versus b)  0.0099 0.0000 0.0008
P (versus c)  0.0000 0.0000
P (versus d)  0.0078

Note: SD represents standard deviation. P values were obtained by the two-tailed Student's t-test, comparing between two types of anchors as indicated.