Skip to main content
. 2012 Jan 11;12:2. doi: 10.1186/1472-6904-12-2

Table 2.

Validation results

Analyte Measuring Range (ng/ml)
MPA: μg/ml
n = 3
LOQ, CV (%)
Accuracy
(%)
n = 5
Sample Conc.
(ng/ml)
MPA: μg/ml
Intra Assay Precision
(%)
n = 5
Inter Assay Precision (%)
n = 20
Accuracy (%)
Intra/Inter
Assay
Cyclosporine A 2 - 1250 3.7 2 7.4 7.1 125/113
r2= 0.997 104 100 2.3 2.9 104/106
500 0.9 2.5 106/105

Tacrolimus 0.5 - 42.2 7.5 0.5 12.3 9.1 138/100
r2= 0.998 108 10 3.7 4.9 116/99
20 3.8 4.5 111/98

Sirolimus 0.62 - 49.2 4.6 1 12.0 11.5 125/107
r2= 0.999 106 10 5.1 5.0 100/102
20 5.0 4.9 108/102

Everolimus 0.53 - 40.8 6.4 1 14.7 12.5 89/90
r2= 0.999 118 10 2.9 6.9 97/96
20 5.1 5.7 102/97

MPA 0.01 - 7.5 7.0 0.02 11.0 9.8 97/102
r2= 0.999 105 0.2 1.3 3.2 97/97
5 1.6 3.0 102/99

(LOQ: low limit of quantification, CV: variation coefficient)