Fig. 3. Reproducibility of SiMoA PSA assay.

Fig. 3

Total imprecision was estimated by repeated measurement of a panel of prepared PSA samples over a 10-day period with two runs/day. Variation sources included fiber strips and processing, inter-calibration, and day-to-day reproducibility. The lowest sample was prepared to approximate the LoQ, and the total imprecision obtained was consistent with the LoQ estimate (20% CV at 0.035 pg/mL).