Table 1.
Before discrepant analysis | After discrepant analysis | ||||||||
---|---|---|---|---|---|---|---|---|---|
| |||||||||
Overall | ELVIS | Overall | ELVIS | ||||||
POS | NEG | Total | POS | NEG | Total | ||||
IsoAmp | POS | 309 | 49 i | 358 | IsoAmp | POS | 351 | 7 | 358 |
NEG | 11 ii | 593 | 604 | NEG | 5 | 599 | 604 | ||
Total | 320 | 642 | 962 | Total | 356 | 606 | 962 | ||
| |||||||||
Value (%) | 95% CI | Value (%) | 95% CI | ||||||
| |||||||||
Positive agreement | 96.6 | 94.0 – 98.1 | Positive agreement | 98.6 | 96.8 – 99.4 | ||||
Negative agreement | 92.4 | 90.1 – 94.2 | Negative agreement | 98.8 | 97.6 – 99.4 | ||||
PPV | 86.3 | 82.4 – 89.5 | PPV | 98.0 | 96.0 – 99.0 | ||||
NPV | 98.2 | 96.8 – 99.0 | NPV | 99.2 | 98.1 – 99.6 | ||||
Prevalence | 33.3 | 30.4 – 36.3 | Prevalence | 37.0 | 34.0 – 40.1 | ||||
Total Agreement | 93.8 | Total Agreement | 98.8 |
POS = positive (HSV detected); NEG = negative (HSV not detected); PPV = positive predictive value; NPV = negative predictive value; CI = confidence interval
49 samples were tested using bidirectional sequencing analysis. Sequence analysis detected HSV target in 42 of the 49 discordant samples [18 HSV-1 and 24 HSV-2] identified as HSV Positive by the IsoAmp® HSV Assay. Sequence analysis did not detect HSV in seven (7) of the discordant samples.
Eleven (11) samples were tested using bidirectional sequencing analysis. Sequence analysis did not detect HSV target in six (6) of the 11 samples identified as HSV Negative by the IsoAmp® HSV Assay. Sequence analysis did detect HSV in five (5) samples [3 HSV-1 and 2 HSV-2]