Skip to main content
. 2012 May 4;303(1):H126–H132. doi: 10.1152/ajpheart.00560.2011

Table 2.

Calibration factor calculations from the aorta

Specific Aortic Pressure Phases Catheter, mmHg Subharmonic Amplitude, dB Calibration Factor, mmHg/dB
Canine 1
    Peak systolic 74.8 ± 1.2 53.0 ± 1.1
    Minimum diastolic 49.6 ± 0.4 65.5 ± 1.3
    Systemic pulse 25.2 ± 1.0 −12.5 ± 1.0 −2.0
Canine 2
    Peak systolic 81.0 ± 1.1 48.6 ± 0.2
    Minimum diastolic 59.9 ± 1.0 54.1 ± 0.3
    Systemic pulse 21.1 ± 0.3 −5.5 ± 0.1 −3.8
Canine 3
    Peak systolic 76.7 ± 1.8 61.6 ± 2.4
    Minimum diastolic 45.3 ± 3.0 71.6 ± 1.9
    Systemic pulse 31.4 ± 3.5 −10.0 ± 3.1 −3.1
Canine 4
    Peak systolic 79.3 ± 0.6 47.6 ± 0.9
    Minimum diastolic 51.7 ± 0.8 59.3 ± 1.3
    Systemic pulse 27.6 ± 1.0 −11.7 ± 1.6 −2.4
Canine 5
    Peak systolic 122.8 ± 0.8 46.9 ± 1.7
    Minimum diastolic 81.0 ± 1.0 57.3 ± 1.1
    Systemic pulse 41.8 ± 1.3 −10.4 ± 2.0 −4.0

Values are means ± 1 SD. Subharmonic signal amplitudes at optimum incident acoustic power setting are indicated. Systemic pulse values are calculated as the difference between the systolic and diastolic values. The calibration factor values are calculated by dividing the systemic pulse pressures by the systemic pulse subharmonic amplitudes.