Table 1. CNRs and CBRs for restored and unprocessed images. The average values were obtained over 200 microspheres in the brain after detection and manual verification.
Algorithm | Mean Peak Value | Background | CNR | CBR |
---|---|---|---|---|
Unprocessed | 26 | 1.7 | 27 +/− 5 | 14 +/− 2 |
RL | 60 | 1.8 | 53 +/− 19 | 32 +/−11 |
Wiener | 54 | 1.77 | 51 +/− 16 | 29 +/− 9 |
Next-Image | 18 | 0.72 | 39 +/−15 | 23 +/− 9 |