Skip to main content
. 2012 Aug;14(8):710–718. doi: 10.1089/dia.2012.0014

Table 3.

Difference Analysis of YSI Glucose Values and Rate of Change of Sensor Glucose Readings

Relative deviation
Absolute value of relative deviation
Group (mg/dL/min) Number % Group (mg/dL/min) Number %
<−3 10 0.8
−3 to <−2 16 1.3
−2 to <−1 57 4.5
−1 to ≤1 1,093 86.1 (0, 1) 1,093 86.1
1 to ≤2 67 5.3 (1, 2) 124 9.8
2 to ≤3 19 1.5 (2, 3) 35 2.8
>3 7 0.6 >3 17 1.3