Fig. 5.
Aberration measurement on fixed Brainbow mouse hippocampus mounted in Vec-tashield. (a) and Media 6 (9.9MB, AVI) , CFP and endogenous 2PEF image at depths of 10, 100 and 200μm. (b)–(d), aberration coefficient maps for (b), astigmatism (z=5); (c), coma (z=8); (d), spherical aberration (z=11). The three maps share the same colour scale. (e), maps of the total aberration amplitude. Scale bar, 100μm. At each depth, measurement of the 11 aberration maps took about 6 minutes and induced ≈ 7% decrease of the fluorescent signal due to photobleaching.