Published in final edited form as: Circulation. 2006 Sep 11;114(12):1277–1284. doi: 10.1161/CIRCULATIONAHA.105.607655

Table.

Comparison of Device Parameters After MRI With Baseline Values Before MRI*

Difference Immediately After MRI P Difference at Follow-Up (Median 99 Days After MRI) P
P-wave amplitude, mV 0.06±0.07 0.44 −0.14±0.27 0.61
R-wave amplitude, mV −0.09±0.15 0.58 0.96±0.70 0.19
Atrial lead impedance, Ω −2.68±5.01 0.60 −6.17±11.53 0.60
Ventricular lead impedance, Ω −0.49±2.74 0.86 6.15±12.56 0.63
Atrial capture, V 0.02±0.02 0.28 −0.07±0.18 0.69
Ventricular capture, V −0.001±0.013 0.93 −0.01±0.05 0.85
Battery voltage, V −0.004±0.002 0.08 −0.05±0.02 0.06
*

Values are reported as mean difference between immediate or long-term and baseline parameter ±SEM. Student t test was not performed on left ventricular lead thresholds in cardiac resynchronization therapy systems because there were not enough data points for a meaningful comparison; there were no significant changes in individual left ventricular lead thresholds after testing, however.