Mass carbon injected vs mass carbon transmitted to the exit capillary per a peak. An average of mass transmitted over mass injected yields a value of 1.0 +/− 0.04.

Mass Carbon Injected (μg) Mass Carbon Transmitted (μg) Transmitted/Injected
6.22E-01 +/− 1.2E-02 6.57E-01 +/− 5.9E-02 1.06 +/− 0.10
2.09E-01 +/− 4.2E-03 2.10E-01 +/− 1.9E-02 1.01 +/− 0.09
7.21E-02 +/− 1.4E-03 6.85E-02 +/− 6.2E-03 0.95 +/− 0.09
2.42E-02 +/− 4.8E-04 2.26E-02 +/− 2.0E-03 0.93 +/− 0.09
8.16E-03 +/− 1.6E-04 8.80E-03 +/− 7.9E-04 1.08 +/− 0.10