Scan protocols for a 1-year-old-equivalent phantom with dose reduction achieved and scoring of image quality

Phantom diameter (cm) Scan protocol kV mA mAs CTDIw (mGy) Dose reduction, % Image quality scoring
10 1 120 32 128 16.59 0 OK
10 2 120 20 80 10.41 37.3 OK
10 3 120 10 40 5.47 67.0 OK
10 4 100 10 40 3.39 79.6 OK
10 5 80 10 40 1.81 89.1 OK
10 6 70 10 40 1.15 93.1 OK
10 7 60 10 40 0.69 95.8 Not OK