Skip to main content
. 2012 Jul 25;39(8):4943–4958. doi: 10.1118/1.4736809

Table 1.

Measurements of the reconstructed attenuation images from the simulation, with sorted projections and interpolated projections, without and with bilinear filtration.

  RMSE (mg iodine/ml) STD (HU) MTF at 0.1 (mm−1) SE (fraction)
Sorted projections 58.6 130 7.9 0.24
Interpolated projections 2.7 44 6.4 0.14
Bilinear filtration 1.5 21 6.8 0.14