TABLE I.

Quantitative Evaluation of the Proposed Eye Movement Correction Algorithm Using Simulated Data

Simulated eye movement
amplitude
Residual shift
in pixel
(percentage)
Sensitivity Specificity
Overall 0.22 (4.81%) 98.85% 99.43%

Small, outside ONH 0.07 (1.72%) 100% 99.42%
Medium, outside ONH 0.38 (6.62%) 100% 99.48%
Large, outside ONH 0.17(1.81%) 100% 99.45%

Small, inside ONH 0.31 (12.93%) 93.10% 99.44%
Medium inside ONH 0.21 (3.60%) 100% 99.45%
Large, inside ONH 0.21 (2.18%) 100% 99.39%

ONH—optic nerve head, overall—2–10 pixels including both inside and outside ONH region, small—2–4 pixels, medium—5–7 pixels, large—8–10 pixels.