Published in final edited form as: J Lab Autom. 2011 May 16;16(3):197–203. doi: 10.1016/j.jala.2011.02.004

Table 3.

Capabilities of different analytical methods and instrumentation used for quality control of HTS compound libraries.

Analytical Instrument or Method Precipitate Empty Wells Presence of Bubbles Sample Volume MW ID Sample Color Analysis time, 384-well plate
Nephelometry10,11,16 yes no no no no no ~5-minutes
Acoustic Devices1215 * yes * yes no no ~7-minutes*
LC-MS9 no no no no yes yes 3–40 hours
HIAPI-CM yes yes yes yes no yes ~1-minute
*

Not designed to unambiguously detect problem but will flag acoustic anomalies; the effect of suspended solids/colloids are unknown. Acoustic methods require proprietary consumables.

Analysis requires destructive sampling