Skip to main content
. 2012 Jun 21;14(1):38. doi: 10.1186/1532-429X-14-38

Figure 7.

Figure 7

In vivo heating measurement. Inner (red) and outer (blue) temperature measurements of the “hot spot” at the worst-case (14 cm) insertion length, at which the guidewire remained inside the introducer sheath, using a 90° flip angle, SAR, 3.96 W/kg.