Table 1.
Centre 1 | Centre 2 | Mean (centre 1 +2) | Pearson correlation coefficient | P value (paired t-test) | P value (Maloney-Rastogi-test) | |||
StatSensor™ | ||||||||
cSCr | Pre offset | Device A | 0.80 ± 0.34 | 0.89 ± 0.27 | 0.85 ± 0.31 | Inter-device evaluation (Device A vs B) | ||
Device B | 0.81 ± 0.34 | 0.89 ± 0.26 | 0.85 ± 0.30 | Center 1 = 0.97 | Center 1 = 0.404 | Center 1 = 0.480 | ||
Difference between Device A vs B | -0.01 ± 0.09 | 0.00 ± 0.09 | Center 2 = 0.95 | Center 2 = 0.689 | Center 2 = 0.521 | |||
Mean (Device A + B) | 0.80 ± 0.34 | 0.89 ± 0.26 | 0.85 ± 0.31 | |||||
Post offset | Device A | 0.97 ± 0.50 | ||||||
Device B | 0.96 ± 0.50 | |||||||
Laboratory reference | ||||||||
vSCr | 0.96 ± 0.42 | 1.02 ± 0.26 | 0.99 ± 0.35 | Inter-method evaluation (cSCr vs vSCr) | ||||
Difference | ||||||||
cSCr - vSCr | Pre offset | Mean (Device A + B) | -0.16 ± 0.16a | -0.14 ± 0.15b | -0.15 ± 0.15c | < 0.0001a,b,c | < 0.0001a,c; 0.9962b | |
Device A | -0.16 ± 0.16 | 0.93 | < 0.0001 | < 0.0001 | ||||
Device B | -0.15 ± 0.17 | 0.92 | < 0.0001 | < 0.0001 | ||||
Post offset | Device A | 0.01 ± 0.19 | 0.93 | 0.638 | < 0.0001 | |||
Device B | 0.00 ± 0.20 | 0.92 | 0.955 | < 0.0001 |
cSCr: Capillary serum creatinine; vSCr: Venous serum creatinine.