. 2012 Jul 28;4(7):328–334. doi: 10.4329/wjr.v4.i7.328

Table 1.

Means, standard deviations, calculated difference between the measured serum creatinine level (mg/dL) using StatSensor™ creatinine and the laboratory reference, Pearson's correlation coefficients, paired t-test and Maloney-Rastogi test, subdivided by study centre, pre- and post-offset correction

Centre 1 Centre 2 Mean (centre 1 +2) Pearson correlation coefficient P value (paired t-test) P value (Maloney-Rastogi-test)
StatSensor™
cSCr Pre offset Device A 0.80 ± 0.34 0.89 ± 0.27 0.85 ± 0.31 Inter-device evaluation (Device A vs B)
Device B 0.81 ± 0.34 0.89 ± 0.26 0.85 ± 0.30 Center 1 = 0.97 Center 1 = 0.404 Center 1 = 0.480
Difference between Device A vs B -0.01 ± 0.09 0.00 ± 0.09 Center 2 = 0.95 Center 2 = 0.689 Center 2 = 0.521
Mean (Device A + B) 0.80 ± 0.34 0.89 ± 0.26 0.85 ± 0.31
Post offset Device A 0.97 ± 0.50
Device B 0.96 ± 0.50
Laboratory reference
vSCr 0.96 ± 0.42 1.02 ± 0.26 0.99 ± 0.35 Inter-method evaluation (cSCr vs vSCr)
Difference
cSCr - vSCr Pre offset Mean (Device A + B) -0.16 ± 0.16a -0.14 ± 0.15b -0.15 ± 0.15c < 0.0001a,b,c < 0.0001a,c; 0.9962b
Device A -0.16 ± 0.16 0.93 < 0.0001 < 0.0001
Device B -0.15 ± 0.17 0.92 < 0.0001 < 0.0001
Post offset Device A 0.01 ± 0.19 0.93 0.638 < 0.0001
Device B 0.00 ± 0.20 0.92 0.955 < 0.0001

cSCr: Capillary serum creatinine; vSCr: Venous serum creatinine.