Table 1.

Spectral-domain (SD) optical coherence tomography (OCT) retinal nerve fibre layer (RNFL) thickness measurement reproducibility

Authors SD-OCT device (company) Study design Main finding
Leung et al18 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) Intravisit repeatability of Cirrus OCT ranged between 5.12 and 15.02 μm, intervisit reproducibility ranged between 4.31 and 22.01 μm
The intervisit variabilities of sectoral and mean RNFL thicknesses were lower in Cirrus HD-OCT compared with TD-OCT (Stratus)
Schuman19 Cirrus OCT (Carl Zeiss Meditec)
RTVue OCT (Optovue)
Comparison with TD-OCT (Stratus) SD-OCT had a significantly better repro- ducibility in most RNFL sectoral measurements than TD-OCT
No statistically significant differences in overall mean RNFL reproducibility between SD- and TD-OCT
Vizzeri et al21 Cirrus OCT (Carl Zeiss Meditec) Reproducibility in healthy and glaucomatous eyes The CV and ICC for mean RNFL thickness were 1.5% and 0.96 in healthy eyes, and 1.6% and 0.98, in patient eyes
González-García et al22 RTVue OCT (Optovue) Reproducibility in healthy and glaucomatous eyes The CV and ICC for mean RNFL thickness were 1.54% and 0.97 in healthy eyes, and 1.9% and 0.97 in patient eyes
Garas et al23 RTVue OCT (Optovue) Influence of various factors on reproducibility The intrasession CV and ICC for mean RNFL thickness were 2.2% and 0.99 in undilated status
Pupil dilation, age and experience in imaging examinations did not influence reproducibility in a clinically significant manner
Menke et al24 3D OCT1000 (Topcon) Intrasession, interobserver reproducibility in healthy eyes The intrasession CV and ICC were 4% (operator 1), 4.2% (operator 2) and 0.90 3D-OCT RNFL thickness measurements in healthy volunteers showed good intra- and interobserver reproducibility
Shin et al25 RTVue OCT (Optovue) Reproducibility of two different scan modes, NHM4 and RNFL 3.45 The CV and ICC for mean RNFL thickness were 2.31% and 0.96 in NHM4 mode, 2.03 and 0.95 in RNFL 3.45 modes
Both NHM4 and RNFL3.45 modes showed excellent measurement reproducibilities

CV, coefficient of variance; HD, high definition; ICC, intraclass correlation coefficient; NHM, nerve head map; TD, time domain.