Table 2.

Spectral-domain (SD) optical coherence tomography (OCT) retinal nerve fibre layer (RNFL) thickness measurement diagnostic capabilities

Authors SD-OCT device (company) Study design Main findings
Leung et al18 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) The AUC was 0.962 for SD- and 0.956 for TD-OCT
No significant difference was detected between the TD- and SD-OCT
Park et al31 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) The AUC was 0.962 for SD- and 0.956 for TD-OCT
SD-OCT showed a better glaucoma discrimination capability than TD-OCT in the early stages of glaucoma
Moreno-Montañés et al32 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) The AUC was 0.837 for SD- and 0.829 for TD-OCT
The sensitivity and specificity, and AUCs were similar between TD and SD-OCT
Cho et al33 SD-SLO/OCT (OTI) Comparison with TD-OCT (Stratus) The AUC was 0.969 for SD- and 0.959 for TD-OCT
Li et al34 RTVue OCT (Optovue) RNFL and ONH parameters The AUC for RNFL thickness was 0.816, and for the cup–disc vertical ratio 0.782
Sehi et al35 RTVue OCT (Optovue) Comparison with TD-OCT (Stratus) The AUC was 0.88 for SD- and 0.87 for TD-OCT
No significant difference was detected between the TD- and SD-OCT
Leung et al36 Spectralis OCT (Heidelberg Engineering) Comparison with HRT (Heidelberg Engineering) SD-OCT RNFL measurement attained a higher sensitivity (AUC; 0.978) than HRT optic disc measurement (0.905)
Jeoung et al37 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) No significant differences between the AUCs for SD- (0.728) and TD-OCT (0.760) in discriminating preperimetric glaucoma
Sung et al29 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) SD-OCT demonstrated a higher sensitivity (63.6%) than TD-OCT (40.0%) in normative classification of mean RNFL thickness
Chang et al30 Cirrus OCT (Carl Zeiss Meditec) Comparison with TD-OCT (Stratus) The sensitivity and specificity of various RNFL parameters using the Cirrus OCT for glaucoma with early to moderate visual field defects are excellent and are equivalent to Stratus OCT
Vizzeri et al28 Cirrus OCT (Carl Zeiss Meditec) Spectralis OCT (Heidelberg Engineering), RTVue OCT (Optovue) Detection of localised RNFL defect All three SD-OCTs were able to detect localized glaucomatous structural damage seen in stereophotographs

AUC, area under the receiver operating characteristic curve; HRT, Heidelberg retinal tomogaph; ONH, optic nerve head; TD, time domain.