Published in final edited form as: Appl Radiat Isot. 2011 Dec 29;70(4):743–751. doi: 10.1016/j.apradiso.2011.12.032

Table 2.

Comparison of activities of point radiation sources: reference values and those calculated with the detector model from background subtracted measurements of exposure rate.

Isotope Reference activity—R (kBq) d = 0.15 cm
d = 10 cm
Backgrounda subtracted measurement of exposure rate (μR h−1) Calculated activity—C (kBq) Relative error (C–R)/R (%) Backgroundb subtracted measurement of exposure rate (μR h−1) Calculated activity—C (kBq) Relative error (C–R)/R (%)
241Am 108.4 1188 122.4 12.9 44.5 121.2 11.8
57Co 41.0 1138 40.6 −1.0 46.5 36.8 −10.2
133Ba 53.7 2138 56.5 5.2 102.5 58.3 8.6
139Ce 16.7 508 16.8 0.6 23.5 17.0 1.8
137Cs 126.9 1238 122.2 −3.7 63.5 112.3 −11.5
228Th 35.1 1088 31.0 −11.7 52.5 30.8 −12.3
152Eu 45.2 1538 44.2 −2.2 70.5 42.6 −5.8
a

Background exposure rate was equal to 12 μR h−1.

b

Background exposure rate was equal to 12.5 μR h−1.