Skip to main content
. 2012 Aug 20;5:39–49. doi: 10.2147/JAA.S33688

Table 1.

Mean injection time into air (seconds) and mean derived volume dispensed (mg) at base conditions for each AAI

AAI type Mean SD 95% CI
Injection time (seconds)
Jext 0.17 0.006 [0.17–0.17]
EpiPen 0.19 0.017 [0.18–0.19]
Anapen 1.16 0.089 [1.13–1.20]
Volume dispensed (mg)
Jext 0.31 0.008 [0.30–0.31]
EpiPen 0.30 0.016 [0.29–0.30]
Anapen 0.29 0.023 [0.28–0.29]

Notes: Jext displayed the shortest injection time and the greatest volume dispensed, followed by EpiPen and Anapen. 15 AAIs of each type were tested.

Abbreviations: SD, standard deviation; CI, confidence intervals; AAI, adrenaline auto-injector.