. 2012 Jul 4;13(7):8324–8337. doi: 10.3390/ijms13078324

Table 1.

Repeatability, reproducibility and accuracy of HPLC method used for aflatoxin determination.

Aflatoxin LOD a (ng g−1) LOQ b (ng g−1) Calibration Curve R2 Recovery (%) c Mean (μg kg−1) ± RSD (%) d
AFB1 0.02 0.05 y = 68,983x + 34,942 0.9997 97.6 125.3 ± 9.12
AFB2 0.01 0.02 y = 104,767x – 6,094 0.9995 91.2 15.3 ± 2.01
AFG1 0.02 0.05 y = 32,045x + 2,780 0.9996 97.6 15.3 ± 1.44
AFG2 0.01 0.02 y = 61,801x − 85.618 0.9991 91.2 6.3 ± 3.42
a

Limit of detection;

b

Limit of quantification;

c

Accuracy was determined by the determination of the recoveries of aflatoxins. By spiking 125.5 μg kg−1 aflatoxin B1, 15.3 μg kg−1 aflatoxin G1 and B2 and 6.3 μg kg−1 G2 to the samples (uninfected ground and tree nuts);

d

Replicate analysis of each spiked sample was used to determine the accuracy, expressed as mean (μg kg−1) ± relative standard deviation (%).