Table 1.

MRI pulse sequence protocol used during the study period: 1.5 T MRI, dedicated torso coil

Pulse sequence Plane TR TE Flip FOV (mm) Gap (mm) Slice (mm) Matrix
SURVEY insp Axial 2.5 1.27 50 450 3.5 8 192 × 144
SURVEY exp Axial 2.5 1.27 50 450 3.5 8 192 × 144
Refscan Axial 8.0 0.57 56 × 40
T1 TFE bh, insp Axial 8.5 4.2 10 450 0 10 256 × 128
T1 TFE bh, insp Sagittal 8.5 4.2 10 450 0 10 256 × 128
T1 TFE bh, insp Coronal 8.5 4.2 10 450 0 10 256 × 128
T1 TFE bh in + out of phase Axial 181 2.3/4.6 80 375 1 7 224 × 134
T2 TSE RT Axial 556 80 90 405 1 7 400 × 215
T1 FFE RT Axial 10 4.6 15 405 1 7 256 × 126
T2 TSE SSH Axial 8,000 800 90 300 40 320 × 256
EPI-DWI b = 0, 50 fb, RT Axial 4,095 56 85 360 0 5 128 × 83
EPI-DWI b = 0, 500 fb, RT Axial 4,095 56 85 360 0 5 128 × 83
THRIVE bh (pre-contrast; 25 and 60 s; 3, 5, 10 and 20 min)a Axial 3.7 1.76 10 450 −2 4 176 × 124

TR Repetition time, TE echo time, flip flip angle, FOV field of view, slice slice thickness, exp expiratory, insp inspiratory, TFE turbo field echo, TSE turbo spin echo, FFE fast field echo, EPI echo planar imaging, SSH single shot, RT respiratory triggered, bh breath hold, fb free breathe, THRIVE T1-weighted high resolution isotropic volume examination

aAfter injection of Gd-EOB-DTPA-DTPA 0.25 μmol/kg at 2 ml/s