Published in final edited form as: Magn Reson Med. 2012 Nov;68(5):1376–1382. doi: 10.1002/mrm.24132

Table 1.

Comparison ChoCre/Cit measurement for PSF-Choice and standard phase encoding using the diagnostic-quality phantom. Error magnitude was computed as magnitude of the difference between Normal compartment measurements (N=20) and Cancer compartment measurements (N=58) and the true values of 0.70 and 1.43 respectively.

Mean and Standard
Deviation (sd) of
ChoCre/Cit Ratio
Normal
Compartment
True Value=0.70
sd = 0.07
Cancer
Compartment
True Value=1.43
sd = 0.18
Mean of Error
Magnitude in
Normal
Compartment
Measurement
Mean of Error
Magnitude in
Cancer
Compartment
Measurement
PSF-Choice 0.74
sd = 0.15
1.36
sd = 0.38
0.14 0.29
Standard Method 0.87
sd = 0.20
1.12
sd = 0.43
0.20 0.47
Error (PSF-Choice) < Error (Standard Method) → P=0.2 P=0.0002