Technical characteristics of standard CMR sequences

Sequence Pixel size (mm) Read FOV 340 mm Pixel size (mm) Phase FOV 340 mm Flip angle (degrees) TE (ms) TR (ms) TI (ms) Interslice gap (mm) Slice thickness (mm) Raw data (lines/cycle or turbo factor)
Breath-hold balanced SSFP cine
1.8
1.8
80
1.3
40(2.6)
 
3
7
15
T2W-TSE with dark blood preparation and fat-suppression by STIR (short tau inversion recovery)
1.3
2.3
 
47
Alternative cardiac cycles
170
3
7
15
Early Gd enhancement (inversion-recovery spoiled gradient-echo)
1.3
1.9
40
1
Alternative cardiac cycles
400
3
7
25
Late Gd enhancement (inversion-recovery spoiled gradient-echo) 1.3 1.9 40 1 Alternative cardiac cycles 260-400 3 7 25