Diabetes is a disease of numbers. Patients are asked to monitor blood glucose (BG) levels, insulin doses, exercise, calories, blood pressure, and severity of symptoms on numerical scales. With so many numbers to keep track of, diabetes patients can benefit from monitoring devices to measure, store, and analyze data.1 New sensors, software, communication technologies, and motivational methods are elevating real-time monitored data to a position of increasing importance as a key component of diabetes management.
Principles of Monitoring for Diabetes
Current and emerging technniques for monitoring diabetes are based on four principles; these are listed in Table 1. Diabetes monitoring diabetes leads to better outcomes, provided that patients demonstrate adequate adherence, use the monitoring to make therapy adjustments, receive adequate education on how to respond to the measurements, and utilize the emerging power of telemedicine support that is increasingly associated with medical monitors.
Table 1.
Principles of Monitoring Patients with Diabetes
| Monitoring can only improve outcomes if patients actually use the monitors as prescribed |
| The purpose of real-time monitoring is to adjust treatment in real time—not just to enter data into a log |
| Health care professionals must educate their patients about what to do with the monitored data |
| Smart monitors with telemedicine can now analyze data, recom-mend therapy, or send alerts |
Monitoring of diabetes is currently being performed on at least six types of health measures; they are listed in Table 2. Furthermore, new sensors for physiological monitoring are currently being developed, and the list will surely expand in the next few years. This editorial focuses on monitoring of BG, continuous glucose, and closed-loop system performance.
Table 2.
Health Monitoring Currently Being Performed by Patients with Diabetes on Themselves
| 1. BG |
| 2. Continuous glucose |
| 3. Closed-loop system performance |
| 4. Adherence to oral medications |
| 5. Blood pressure |
| 6. Foot images |
| 7. Exercise performed |
| 8. Dietary intake |
Blood Glucose Monitoring
Self-monitoring of blood glucose (SMBG) has four main purposes. Purposes of this practice include (1) protecting by allowing immediate confirmation of low or high BG levels; (2) self-treating high or low BG levels by adjusting therapy to reach target glycated hemoglobin (HbA1c); (3) educating about the effects of diet, exercise, and other factors on BG levels; and (4) motivating healthy behavior. Methods for data management of SMBG data are listed in Table 3.
Table 3.
Methods for Data Management of Self-Monitored Blood Glucose Data
| Logbook |
| Web-based software |
| Smart phone via cable |
| Wireless to central server and then to multiple communication sys-tems, including smart phone, telephone, or fax |
Data recorded by hand in logbooks are subject to reporting errors. In one study in France, subjects' BG values were recorded and stored by the BG monitors without their knowledge. Overreporting (by writing phantom values in a logbook) and underreporting (by omitting SMBG measurements from a written logbook) occurred for 19% and 12%, respectively, of the actual data points.2 Patients frequently report during visits that they forgot to bring their logbooks or their monitors that contain a log of BG data. Many health care professionals assume that, in most of these cases, the patients did not actually test themselves as much as was requested, but this situation has not been formally studied. Only a technology that can automatically time stamp, store, and transmit SMBG data to a health care professional can eliminate the problem of forgotten logbooks and meters and deliver accurate monitoring data. Furthermore, for the stored and transmitted data to be useful, the system must analyze the data stream with pictorial graphical displays and calculated metrics.
Manufacturers of most BG monitors provide dedicated software for analyzing data from their monitor, which can be downloaded from the Internet onto a computer. Data is then uploaded into the software program either by cable or wirelessly. The analyzed data can then be printed and brought to the clinician's office.3 An emerging method for analyzing BG data is to transmit the data from the glucose monitor to a server and back to the person's smart phone for data analysis and, in some cases, decision support. Data can be transmitted by cable from the monitor to the smart phone for automatic uploading or from the monitor directly into the smart phone (if there is a connection between the two devices) or by direct transmission from the monitor to a central server (without a cable or a smart phone). From the server, the information can be sent automatically to a computer or a smart phone. This process of automatic uploading is expected to become increasingly adopted even as many other numbers that are important in our lives are becoming stored on the Internet.4
Software Applications for Blood Glucose Data
Mobile software applications, known as apps, for analyzing BG data and other numerical measures of interest to patients with diabetes are becoming increasingly available. Mobile health care and medical app downloads globally are expected to reach 44 million users by the end of 2012 and 142 million users by 2016.5 A survey of 21 patients with type 1 diabetes who had not previously worked with any diabetes apps determined that the seven most important features in a SMBG app, starting with the most important, were (1) ease of use; (2) communication with the meter; (3) wireless connectivity; (4) number of reports; (5) appearance of the screens; (6) price; and (7) ability to synchronize with an online account.6
A literature review of features of SMBG mobile diabetes applications ranked the most common features that accompanied display and analysis of SMBG data.7 The six most common features are displayed according to rank and prevalence in Table 4.
Table 4.
Six Most Prevalent Features in Diabetes Apps in Addition to Self-Monitoring of Blood Glucose Data Analysis, According to Rank and Prevalence7
| Rank | App feature | Prevalence |
|---|---|---|
| 1 | Insulin dosage | 65% |
| 2 | Communication | 59% |
| 3 | Diet | 52% |
| 4 | Physical activity | 40% |
| 5 | Weight | 39% |
| 6 | Blood pressure | 32% |
Peripherals for Monitoring of Diabetes
The capabilities of smart phones are being extended by the attachment of peripheral equipment, or peripherals, that contain sensors.8 Many of these sensors are wearable.9 At this time, peripherals are not frequently used, but as the monitoring technology improves, it is expected that smart phones will be used as part of monitoring systems. An example of this process would be the use of a light and fluorometer to read the fluorescence lifetime of nano-optodes, which fluoresce according to the concentration of glucose in the ambient interstitial fluid.10 The light and fluorometer could be a peripheral for a smart phone.11 A touch screen has been developed to be a biomolecule detection platform.12 This system is not currently intended for measuring glucose, but it is conceivable that a sheer disposable screen could be developed to fit over a smart phone and serve as a surface for measuring blood, urine, or saliva specimens to monitor analytes of interest in diabetes.13,14
Adherence to Monitoring
The use of SMBG in non-insulin-treated patients with type 2 diabetes is a controversial topic because some studies have demonstrated improved outcomes and some studies have failed to demonstrate improved outcomes.15 The report of an expert panel that reviewed the literature and reported its findings in 2011 concluded that SMBG in this population should be performed in a structured format where the glucose data are used to guide treatment. New data from randomized controlled trials since 2008 have demonstrated the efficacy of this practice.16 The need for using self-monitoring to guide treatment is a basic feature of all useful types of self-monitoring. Adherence to monitoring will be poor unless patients understand the reasons for monitoring and are taught what to do with the monitored information. The health care professional is obligated to teach the patient what to do with the information, or this tool will probably not be used properly and the monitoring will probably not achieve the intended outcome.
Continuous Glucose Monitoring
The American Association of Clinical Endocrinologists Consensus Panel on continuous glucose monitoring (CGM) reported their conclusions about the use of CGM. The key takeaway or summary of their conclusions for adults was, “More consistent CGM use predicts HbA1c reductions,” and their key takeaway for youth was, “The best HbA1c-lowering results were seen in patients who used the sensor 6 to 7 days a week.”17 These conclusions were based on observations that, in intention-to-treat studies, subjects who use CGM more frequently have better outcomes than subjects who use CGM less frequently.
The STAR 1 trial evaluated the safety and efficacy of an insulin pump combined with use of a real-time CGM device compared with an insulin pump alone in subjects with type 1 diabetes already using insulin pump therapy. The HbA1c reduction was no different between the two groups. Subjects who used CGM during at least 60% of days had a significantly better HbA1c outcome than those who used CGM no more than 60% of days.18 The landmark Juvenile Diabetes Research Foundation CGM study stratified type 1 subjects into three age groups: 25 years or greater, 15–24 years, or 8–14 years. The study demonstrated a benefit of CGM of 0.053%, a worsening of HbA1c of 0.08%, and a benefit of 0.013%, respectively. The use of CGM at least 6 days per week occurred in 83% of subjects ages 25 years or greater, 30% of subjects ages 15–24 years, and 50% of subjects ages 8–14 years. The authors concluded that the observed age effect on HbA1c might have been “related to substantially greater use of sensors in the adults than in patients in the two younger age groups.”19 Vigersky and colleagues20 reported a randomized controlled trial of CGM in type 2 diabetes subjects. The study compared the effects of 12 weeks of intermittent real-time CGM with SMBG after a 40-week follow-up period. The subjects who used the CGM system per protocol (≥48 days) improved the most (p < .0001). These three studies of CGM demonstrated that, when CGM is actually used, the best outcomes occur.
Closed-Loop Systems
Closed-loop control studies have been performed mostly in inpatient studies to date. New trials will be getting underway in the United States and Europe to assess the performance of these systems in outpatients.21 Modular architecture technology is being developed to monitor CGM results and insulin dosages of study subjects remotely.22 Telemedicine systems are being developed to monitor various measures continuously and to contact study subjects immediately if necessary.23 On July 22, 2008, at a Food and Drug Administration (FDA)/National Institutes of Health/Juvenile Diabetes Research Foundation workshop on artificial pancreas in Bethesda, I proposed a remote system to diagnose and fix problems of a closed-loop system. The system was named the Knowledge of Loop Operations Necessary System to Assist Repairs, whose acronym was Klonstar.24 Dassau and associates25 later proposed a specific tele-medicine system based on Klonstar that was called an enhanced 911/global positioning system (GPS) wizard to prevent hypoglycemia, which integrated CGM with a hypoglycemia-predictive algorithm, with a GPS position locator and a short message service to monitor and alert for hypoglycemia. Furthermore, this system would be integrated with the patient's insulin pump to decrease or suspend insulin infusion in the event of impending hypoglycemia.25 Currently, no closed-loop system is approved for use in the United States by the FDA. The first system with automatic control of insulin delivery, Veo, manufactured by Medtronic Diabetes, Northridge, CA, is available in Europe and Canada but has not yet been approved for use in the United States. This system contains hypoglycemic alarms for predicted and threshold levels of hypoglycemia but no remote telemedicine capabilities.26 A remote warning system for use by CGM device users who might be experiencing unrecognized nocturnal hypoglycemia was approved by the FDA. The system consists of a monitor for the patient, a power supply, and a transmitter to send information to a second monitor at the bedside of another person in the house. The second monitor sounds a loud alarm in the event of hypoglycemia, which is intended to awaken the other person in case the patient develops nocturnal hypoglycemia and sleeps through the alarms.27
Conclusions
Many additional monitors for measuring, storing, and assessing information about new analytes and vital signs of interest for diabetes are expected to become available.28–30 Patterns of glycemia over geographic space will be available with the use of GPS combined with CGM.31 Reminders for overdue health maintenance services will be programmable into monitors used at home.4 Increasing use of decision support software for real-time therapeutic maneuvers will add value to monitoring systems that accrue and analyze data.32 Monitoring of BG and many other factors affecting the health of patients with diabetes will be successfully adopted for the benefit of patients (1) if the monitoring equipment is actually used by patients to make measurements; (2) if the information provided from these monitors is actually used by patients to determine therapy; (3) if the use of this equipment is supported by specific educational initiatives from health care providers; and (4) if monitoring systems can be integrated with online Web sites to store and analyze the monitored data. Monitoring for diabetes is in its infancy. This activity will increase in importance in parallel with anticipated advances in clinical chemistry, telemetry, and the pathophysiology of diabetes.
Glossary
Abbreviations
- (HbA1c)
glycated hemoglobin
- (BG)
blood glucose
- (CGM)
continuous glucose monitoring
- (FDA)
Food and Drug Administration
- (GPS)
global positioning system
- (SMBG)
self-monitoring of blood glucose
References
- 1.Kerr D. Poor numeracy: the elephant in the diabetes technology room. J Diabetes Sci Technol. 2010;4(6):1284–1287. doi: 10.1177/193229681000400601. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ziegler O, Kolopp M, Got I, Genton P, Debry G, Drouin P. Reliability of self-monitoring of blood glucose by CSII-treated patients with type I diabetes. Diabetes Care. 1989;12(3):184–188. doi: 10.2337/diacare.12.3.184. [DOI] [PubMed] [Google Scholar]
- 3.Hirsch IB. Blood glucose monitoring technology: translating data into practice. Endocr Pract. 2004;10(1):67–76. doi: 10.4158/EP.10.1.67. [DOI] [PubMed] [Google Scholar]
- 4.Ciemins E, Coon P, Sorli C. An analysis of data management tools for diabetes self-management: can smart phone technology keep up? J Diabetes Sci Technol. 2010;4(4):958–960. doi: 10.1177/193229681000400427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Cox A. Press release: mobile healthcare and medical app downloads to reach 44 million next year, rising to 142 million in 2016. JuniperResearch. http://www.juniperresearch.com/viewpressrelease.php?id=352&pr=275. Accessed March 3, 2012.
- 6.Rao A, Hou P, Golnik T, Flaherty J, Vu S. Evolution of data management tools for managing self-monitoring of blood glucose results: a survey of iPhone applications. J Diabetes Sci Technol. 2010;4(4):949–957. doi: 10.1177/193229681000400426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Chomutare T, Fernandez-Luque L, Arsand E, Hartvigsen G. Features of mobile diabetes applications: review of the literature and analysis of current applications compared against evidence-based guidelines. J Med Internet Res. 2011;13(3):e65. doi: 10.2196/jmir.1874. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Matlack K. 5 medical peripherals for the iPad or iPhone. http://blog.softwareadvice.com/articles/medical/5-medical-peripherals-for-the-ipad-or-iphone-1012612/. Accessed March 3, 2012.
- 9.Chen M, Gonzalez S, Vasilakos A, Cao H, Leung VC. Body area networks: a survey. Mobile Network Applications. 2011;16:171–193. [Google Scholar]
- 10.Balaconis MK, Billingsley K, Dubach MJ, Cash KJ, Clark HA. The design and development of fluorescent nano-optodes for in vivo glucose monitoring. J Diabetes Sci Technol. 2011;5(1):68–75. doi: 10.1177/193229681100500110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Vezina K. Tattoo tracks sodium and glucose via an iPhone. TechnologyReview. http://www.technologyreview.com/computing/38065/. Accessed March 3, 2012.
- 12.Won BY, Park HG. A touchscreen as a biomolecule detection platform. Angew Chem Int Ed Engl. 2012;51(3):748–751. doi: 10.1002/anie.201105986. [DOI] [PubMed] [Google Scholar]
- 13.Ross V. More fun than a blood test: researchers want diagnosis to be as simple as spitting on your screen. Discover Magazine. http://blogs.discovermagazine.com/80beats/2011/12/05/more-fun-than-a-blood-test-researchers-want-diagnosis-to-be-as-simple-as-spitting-on-your-screen/. Accessed March 3, 2012. [Google Scholar]
- 14.Heidesch T. Would you pee on your smartphone? Advance. http://community.advanceweb.com/blogs/nppa_2/archive/2012/02/22/would-you-pee-on-your-smartphone.aspx. Accessed March 3, 2012.
- 15.Klonoff DC, Bergenstal R, Blonde L, Boren SA, Church TS, Gaffaney J, Jovanovic L, Kendall DM, Kollman C, Kovatchev BP, Leippert C, Owens DR, Polonsky WH, Reach G, Renard E, Riddell MC, Rubin RR, Schnell O, Siminiero LM, Vigersky RA, Wilson DM, Wollitzer AO. Consensus report of the coalition for clinical research-self-monitoring of blood glucose. J Diabetes Sci Technol. 2008;2(6):1030–1053. doi: 10.1177/193229680800200612. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Klonoff DC, Blonde L, Cembrowski G, Chacra AR, Charpentier G, Colagiuri S, Dailey G, Gabbay RA, Heinemann L, Kerr D, Nicolucci A, Polonsky W, Schnell O, Vigersky R, Yale JF. Consensus report: the current role of self-monitoring of blood glucose in non-insulin-treated type 2 diabetes. J Diabetes Sci Technol. 2011;5(6):1529–1548. doi: 10.1177/193229681100500630. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Blevins TC, Bode BW, Garg SK, Grunberger G, Hirsch IB, Jovanovič L, Nardacci E, Orzeck EA, Roberts VL, Tamborlane WV, AACE Continuous Glucose Monitoring Task Force, Rothermel C Statement by the American Association of Clinical Endocrinologists Consensus Panel on continuous glucose monitoring. Endocr Pract. 2010;16(5):730–745. doi: 10.4158/EP.16.5.730. [DOI] [PubMed] [Google Scholar]
- 18.Hirsch IB, Abelseth J, Bode BW, Fischer JS, Kaufman FR, Mastrototaro J, Parkin CG, Wolpert HA, Buckingham BA. Sensor-augmented insulin pump therapy: results of the first randomized treat-to-target study. Diabetes Technol Ther. 2008;10(5):377–383. doi: 10.1089/dia.2008.0068. [DOI] [PubMed] [Google Scholar]
- 19.Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group. Tamborlane WV, Beck RW, Bode BW, Buckingham B, Chase HP, Clemons R, Fiallo-Scharer R, Fox LA, Gilliam LK, Hirsch IB, Huang ES, Kollman C, Kowalski AJ, Laffel L, Lawrence JM, Lee J, Mauras N, O'Grady M, Ruedy KJ, Tansey M, Tsalikian E, Weinzimer S, Wilson DM, Wolpert H, Wysocki T, Xing D. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008;359(14):1464–1476. doi: 10.1056/NEJMoa0805017. [DOI] [PubMed] [Google Scholar]
- 20.Vigersky RA, Fonda SJ, Chellappa M, Walker MS, Ehrhardt NM. Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care. 2012;35(1):32–38. doi: 10.2337/dc11-1438. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Heinemann L, Benesch C, DeVries JH. AP@home: a novel European approach to bring the artificial pancreas home. J Diabetes Sci Technol. 2011;5(6):1363–1372. doi: 10.1177/193229681100500607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Lanzola G, Capozzi D, Serina N, Magni L, Bellazzi R. Bringing the artificial pancreas home: telemedicine aspects. J Diabetes Sci Technol. 2011;5(6):1381–1386. doi: 10.1177/193229681100500609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Rigla M. Smart telemedicine support for continuous glucose monitoring: the embryo of a future global agent for diabetes care. J Diabetes Sci Technol. 2011;5(1):63–67. doi: 10.1177/193229681100500109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Klonoff DC. Compatibility of the artificial pancreas with other devices. Presented at: Artificial Pancreas Workshop. July 22, 2008.
- 25.Dassau E, Jovanovic L, Doyle FJ, 3rd, Zisser HC. Enhanced 911/global position system wizard: a telemedicine application for the prevention of severe hypoglycemia--monitor, alert, and locate. J Diabetes Sci Technol. 2009;3(6):1501–1506. doi: 10.1177/193229680900300632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Agrawal P, Welsh JB, Kannard B, Askari S, Yang Q, Kaufman FR. Usage and effectiveness of the low glucose suspend feature of the Medtronic Paradigm Veo insulin pump. J Diabetes Sci Technol. 2011;5(5):1137–1141. doi: 10.1177/193229681100500514. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Medtronic. Medtronic launches first-of-its-kind mySentry™ remote glucose monitor. http://wwwp.medtronic.com/Newsroom/NewsReleaseDetails.do?itemId=1325625195470&lang=en_US. Accessed February 28, 2012. [Google Scholar]
- 28.Lim H. Smart underwear for diabetic patients. JTATM. 2009;6(1):1–11. [Google Scholar]
- 29.Xiang Y, Lu Y. Using personal glucose meters and functional DNA sensors to quantify a variety of analytical targets. Nat Chem. 2011;3(9):697–703. doi: 10.1038/nchem.1092. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Darwish A, Hassanien AE. Wearable and implantable wireless sensor network solutions for healthcare monitoring. Sensors (Basel) 2011;11(6):5561–5595. doi: 10.3390/s110605561. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Doherty ST. Exploring blood glucose variation over geographical space. Diabetes Technol Ther. 2012;14(3):276–284. doi: 10.1089/dia.2011.0159. [DOI] [PubMed] [Google Scholar]
- 32.Klonoff DC, True MW. The missing element of telemedicine for diabetes: decision support software. J Diabetes Sci Technol. 2009;3(5):996–1001. doi: 10.1177/193229680900300501. [DOI] [PMC free article] [PubMed] [Google Scholar]